Hurotics, a Korean maker of soft wearable robots, says it has shown that the two thigh sensors on its H-Medi gait rehabilitation exosuit are enough to estimate each leg's step length and which side of a patient's gait is out of balance. The study was published in the journal Frontiers in Bioengineering and Biotechnology.
H-Medi helps patients straighten the hip as they walk, and uses the inertial sensors on both thighs to control that help. The company's algorithm uses the same two sensors to analyse walking, without extra sensors on the feet or ankles, and needs only the leg length and a two-second standing reading to calibrate.
The researchers recorded 7,268 gait cycles on a treadmill with the two belts at the same and at different speeds, and compared the estimates with an optical motion capture system. Step length agreed with an intraclass correlation of 0.856 and an error of 6.8 cm, and walking speed with a correlation of up to 0.925.
With the belts at different speeds, the method found the side with the longer step 95.2% of the time and the side with the faster step 98.4% of the time. Across six main gait measures, it named the side of the imbalance correctly 91.0% to 98.4% of the time.
Background
H-Medi
Rehabilitation exosuit
Hurotics' soft exosuit medical device that helps patients straighten the hip as they walk.